I set up the K3 Rx sensitivity test (using 1 microvolt signal) according to
XG2 manual.
It seems to work in general except that the resulting MDS calculations seem
too low:
123 dBm on 80 meters,
132 dBm on 40 meters, and
130 dBm on 20 meters.
I am using the front headphone jack on the K3, with the CONFIG:SPKR = 1. AGC
is off.
Preamp is own, RF & AF gains are at maximum, and the CW passband is set to
500 Hz.
The AC measurements were made using a Craftsman autoranging multimeter.
Example: on 40 meters the reading was 0.310 V with XG2 on and 0.018 V with
XG2 off.
(0.310/0.018) = 17.2.
Log 17.2 x 20 = -24.7 dBm.
MDS = -107 dBm -24.7 dBm = 131.7 dBm.
I was expecting a better value (maybe -137 dBm?)
Could it be a problem with the DMM? If so, is there a recommended DMM?
Thanks in advance for your help.
73, Fred AE6IC