Re: K3 SWR Accuracy - reprise
Posted by
ab2tc on
Nov 05, 2009; 12:23am
URL: http://elecraft.85.s1.nabble.com/K3-SWR-Accuracy-reprise-tp3943810p3949304.html
Hello,
Please, please, does anybody remember the original poster's problem anymore?
Quote:
The results I got for 2:1 VSWR are as follows:-
LP-100 K3
160 1.99 1.8
80 1.99 1.8
60 1.99 1.8
40 1.95 1.8
30 1.97 1.6
20 1.97 1.4
17 1.97 1.3
15 1.96 1.2
12 1.96 1.5
10 1.96 1.4
I am repeating. This makes no sense given that he is using a parallel combination of two dummy loads which results in a 2:1 SWR. I am repeating: Either there is something wrong with the data or the K3 is faulty. Errors this large should not occur with properly functioning equipment.
AB2TC - Knut
Steve Ellington wrote
The first statement is correct. Length of coax will transform impedance and
cause SWR meters to read differently.
<dsnip>